Feeds:
Posts
Comments

Posts Tagged ‘intraoperative radiation’

Live Conference Coverage: International Dialogue in Gynecological Oncology, From Bench to Bedside, Ovarian Cancer

Reporter: Stephen J. Williams, Ph.D.

Join Live on Wednesday May 22, 2024 for an international discussion on the current state of ovarian cancer diagnostics and therapeutics, and potential therapies and biomarkers, and biotargets.  Topics including potential new molecular targets for development of ovarian therapeutics, current changes in ovarian cancer clinical treatment protocols, chemo-resistance, and the use of Artificial Intelligence (AI) in the diagnosis and treatment of cancer will be discussed.

International Dialogue in Gynecological Oncology, From Bench to Bedside, ovarian Cancer meeting flyer

To join by Zoom click the link below

https://temple.zoom.us/j/94458267823 

Agenda:

Introduction

  • 00/15.00 Professor Giordano and Professor Ercoli
  • 10/15.10 We Have Never Been Only Human: a new perspective to defeat ovarian cancer (C. Martinelli)

Molecular Section

  • 20/15.20 DNA Repair mechanisms: understanding their role in cancer development and chemoresistance (L. Alfano)
  • 35/15.35 Progranulins: a new target for oncological treatment (A. Morrione)
  • 50/15.50 Modulation of gene expression and its applications (M. Cuomo)
  • 10.05/16.05 Commanding the cell cycle: the role of CDKs (S.R. Burk
  • 10.20/16.20 Drug development from nature (M. D’Angelo

Clinical Section

  • 05/17.05 Core principles of Radiologic Diagnosis & Staging in Ovarian Cancer(A. Blandino)
  • 20/17.20 Key Indications for Nuclear Medicine in Ovarian Cancer (S. Baldari)
  • 35/17.35 Cutting Edge Decision: Understanding Surgical Indications and Outcomes in Ovarian Cancer (A. Ercoli)
  • 50/17.50 Gold Standard in Oncology for Ovarian Cancer (N. Silvestris)
  • 12.05/18.05 Role of Radiotherapy in Ovarian Cancer (S. Pergolizzi)

Conclusion

12.20/18.20 AI Applied to medical science (V. Carnevale)

Speakers

  • – Professor Alfredo Blandino: Professor Blandino holds the esteemed positions of Head of school of Radiology and director of the department of radiology at the University of Messina. He has made significant contributions to diagnostic imaging with over hundreds of publications to his name, Professor Blandino’s work exemplifies excellence and innovation in radiology.
  • – Professor Alfredo Ercoli, serves as the Director of the Department of Gynecology and Obstetrics at the “G. Martino” University Hospital in Messina. He is also head of school of gynecology and obstetrics at Messina University. Starting his research in France with studies on pelvic anatomy that became a cornerstone in medical literature, He is a pioneer in advanced gynecologic surgery, including laparoscopic and robotic procedures, having performed over thousands of surgical interventions. His research focuses on gynecologic oncology, advanced gynecologic surgery, and endometriosis, urogynecology. Professor Ercoli’s dedication to education and his numerous publications have significantly advanced the field of gynecology.
  • Professor Sergio Baldari, an eminent figure in nuclear medicine. Professor Baldari is the Director of the department of nuclear medicine and head of school of nuclear medicine at the  University of Messina. He has authored or co-authored over 500 publications, with a focus on diagnostic imaging and the use of PET and radiopharmaceuticals in cancer treatment. His leadership and expertise have been recognized through various prestigious positions and awards within the medical community.
  • – Professor Nicola Silvestris is the Director of UOC Oncologia Medica at the University of Messina. His extensive research in cancer, has led to over 360 peer-reviewed publications. Professor Silvestris has made significant contributions to translational research and the development of guidelines for managing complex oncological conditions. His work continues to shape the future of cancer treatment.
  • Professor Stefano Pergolizzi, a leading expert in radiation oncology. Professor Pergolizzi serves as the Director of the department of radiotherapy and head of the school of radiotherapya at the University of Messina. He is also the president of the Italian Association of Radiotherapy and Clinical Oncology (AIRO) His research focuses on advanced radiotherapy techniques for cancer treatment. With a career spanning several decades, Professor Pergolizzi has published numerous papers and has been instrumental in developing innovative therapeutic approaches. His dedication to patient care and education is exemplary.
  • Margherita D’angelo: Graduated in Molecular Biology with honors from the Federico II University of Naples.
    Third year intern in Food Science at the Luigi Vanvitelli University of Naples.
    Research intern in Molecular oncology with the project of developing novel drugs starting from food waste at the Sbarro Institute for Cancer Research and Molecular Medicine at Temple University, Philadelphia (USA), directed by Dr A. Giordano.
  • Vincenzo Carnevale, Ph.D.

Dr. Carnevale is an Associate Professor in the Institute for Computational Molecular Science in the College of Science & Technology, Temple University.  He holds multiple NIH RO1 and NSF grants. Vincenzo Carnevale received B.Sc. and M.Sc. degrees in Physics from the University of Pisa and a PhD from SISSA – Scuola Internazionale Superiore di Studi Avanzati in Trieste, Italy. The Carnevale research group uses statistical physics and machine learning approaches to investigate sequence-structure-function relations in proteins. A central theme of the group’s research is how interactions give rise to collective phenomena and complex emergent behaviors. At the level of genes, the group is interested in epistasis – the complex entanglement phenomenon that causes amino acids to evolve in a concerted fashion – and how this shapes molecular evolution. At the cellular level, the group investigates how intermolecular interactions drive biomolecules toward self-organization and pattern formation. A long-term goal of the group is understanding the molecular underpinnings of electrical signaling in excitable cells. Toward these goals, the group applies and actively develops an extensive arsenal of theoretical and computational approaches including statistical (mean)field theories, Monte Carlo and molecular dynamics simulations, statistical inference of generative models, and deep learning.

  • Professor Andrea Morrione, Ph.D: Research Associate Professor, CST Temple University; After his studies in Biochemistry at Universita’ degli Studi Milano, Milan Italy, Dr. Morrione moved to USA in 1993 and has been working in the field of cancer biology since his postdoctoral training at the Kimmel Cancer Institute, Thomas Jefferson University, Philadelphia, PA in the laboratory of Dr. Renato Baserga, one of the leading experts in IGF-IR oncogenic signaling. In 1997 Dr. Morrione joined the Faculty of Thomas Jefferson University in the Department of Microbiology. In 2002 after receiving an NIH/NIDDK Career Development Award Dr. Morrione joined the Department of Urology at Jefferson where from 2008 to 2018 serves as the Director for Urology Basic Science and Associate Professor. Dr. Morrione joined the Department of Biology and the Sbarro Institute for Cancer Research and Molecular Medicine and Center for Biotechnology as Associate Professor of Research, and he is currently professor of Research and Deputy Director of the Sbarro Institute for Cancer Research and Molecular Medicine and Center for Biotechnology. He is a full member of the AACR.

 

  • Canio Martinelli, M.D.: Dr. Marinelli received his MD from Catholic University of the Sacred Heart in Rome, Visiting researcher at SHRO Temple University in Philadelphia, PhD candidate in Translational Molecular Medicine and Surgery & GYN-OB resident at UNIME. He has published numerous clinical papers in gynecologic oncology, risk reduction, and therapy and, most recently investigating clinical utilities of generative AI in gynecologic oncology.
  • Sharon Burk, Sharon Burk is a PhD student with Professor Antonio Giordano at the University of Siena, Italy in the department of Medical Biotechnologies, studying the role of Cyclin Dependent Kinase 10 in Triple Negative Breast Cancer. She received her Bachelor’s of Arts Degree from the University of California, Berkeley with a double major in molecular and cell biology and Italian studies.   She is a member of AACR.

This conference is being sponsored by Sbarro Health Research Organization and the Department of Biology, College of Science & Technology, Temple University.

To join by Zoom click the link below

https://temple.zoom.us/j/94458267823 

A QR code will be supplied at conference start, in addition to Zoom chat, to allow for questions to be submitted.

This conference is free to join on Zoom and will be covered live on @pharmaBI 

and on

 

Read Full Post »

Low Energy Photon Intra-Operative Radiotherapy System

Larry H. Bernstein, MD, FCAP, Curator

LPBI

 

The Dosimetric Characteristics and Potential Limitation in Clinical Application of a Low Energy Photon Intra-Operative Radiotherapy System

Spring Zhou Editor at Scientific Research Publishing        https://www.linkedin.com/grp/post/143951-6069684489951391748

 

Purpose: To investigate the dosimetric characteristics of a low energy photon intra-operative radiotherapy (IORT) system and explore its potential limitation in clinical application.

Methods: A special water phantom, a parallel-plate ionization chamber and an electrometer were used to measure the depth dose rate, isotropy of dose distribution in X/Y plane, dosimetry reproducibility of bare probe and spherical applicators of different size which were used in comparison with the system data.

Results: The difference in depth dose rate between the measurement and system data for bare probe is -2.16% ± 1.36%, the range of the relative deviation for isotropy in the X/Y plane is between -1.9% and 2.1%. The difference in depth dose rate, transfer coefficient, isotropy in X/Y plane between the measurement and system data for the whole set of spherical applicators is -10.0% – 2.3%, -8.9% – 4.2% and -1.6% – 2.6%, respectively. Higher surface dose rate and steeper gradient depth dose are observed in smaller spherical applicators. The depth dose rate and isotropy for bare probe and spherical applicators have been shown good reproducibility. The uncertainty of measurement is associated with the positioning accuracy, energy response, noise current and correction function f’(R).

Conclusions: Thorough commissioning of the low energy photon IORT system helps us better understand the dosimetry characteristics, verify the system data, obtain adequate data for clinical application and routine quality assurance. The steep gradient depth dose and limited treatment range may restrain its potential in clinical application.

http://www.scirp.org/journal/PaperInformation.aspx?PaperID=56700&amp

 

The Dosimetric Characteristics and Potential Limitation in Clinical Application of a Low Energy Photon Intra-Operative Radiotherapy System

IJMPCERO> Vol.4 No.2, May 2015    DOI: 10.4236/ijmpcero.2015.42023

Zhenhua Xiao, Ouyang Bin, Zhenyu Wang, Botian Huang, Bixiu Wen*

 

Cite this paper

Xiao, Z. , Bin, O. , Wang, Z. , Huang, B. and Wen, B. (2015) The Dosimetric Characteristics and Potential Limitation in Clinical Application of a Low Energy Photon Intra-Operative Radiotherapy System. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 4, 184-195. doi:10.4236/ijmpcero.2015.42023.
Intra-operative radiotherapy (IORT) delivers single high dose radiation directly to the tumor bed within a relative short period of treatment time during the surgical operation, which requires a higher surface dose to protect the deep normal tissues. It often uses high energy electron beam or low energy photon beam. Modern IORT is usually delivered in the operating room, which requires that the device is light weighted and easy to move with high quality of radiation protection. Dedicated IORT device commercially available includes Mobetron® MeV electron beam system [1] (Intra Op Medical Corporation, California, USA), NOVAC™ 7 system (New Radiant Technology SpA, Italy) and the INTRABEAM® 50 kV X-ray device (Carl Zeiss Medical Company, Germany).
Since the results from the randomized TARGIT A trial were published, IORT has been applied for locally advanced or recurrent rectal cancer [2], superficial cutaneous malignancies [3] and as full dose partial breast irradiation (PBI) or as a boost after whole-breast radiatherapy (WBI) in early stage breast cancer [4] [5]. Vaidya J.S. et al. have analysis overall survival of using as single-dose targeted intraoperative radiotherpay (TARGIT) versus fractionated external beam radiotherapy (EBRT) for breast cancer. For patients enrolled at 33 centers in 11 countries, 1721 patients were randomised to TARGIT and 1730 to EBRT. The 5-year risk for local recurrence in the conserved breast was 3.3% (95% CI 2.1 – 5.1) for TARGIT versus 1.3% (0.7 – 2.5) for EBRT (p = 0.042). Wound-related complications were much the same between groups but grade 3 or 4 skin complications were significantly reduced with TARGIT (four of 1720 vs 13 of 1731, p = 0.029) [6]. TARGIT-B (for boost, ISRCTN43138042) is an ongoing multicenter randomised controlled trial that began in 2013, which is testing the replacement of EBRT boost to the tumor bed by a TARGIT boost given during surgery. An ongoing open registry study, TARGIT-R (for registry, ISRCTN91179875), began in 2013, aiming to monitor the long-term effectiveness and safety of the patients treated with TARGIT following breast conserving surgery for early breast cancer [7].
Eaton D.J. et al. described the dosimetry of the INTRABEAM® system with the spherical applicators [8]. Maxime Goubert et al. have reported dosimetric behavior of the system equipped with flat and surface applicators [3]. There still remains controversary for the dosimetric characteristics and clinical application of IORT. In this paper we introduce the commissioning test of the INTRABEAM® 50 kV x-ray system by investigating the dosimetric characteristics of the low energy photon IORT system and explore its potential limitation in clinical application.
……
The device for measurement includes a dedicated water tank (see in Figure 1) with radiation protection design (Carl Zeiss Surgical GmbH), a parallel plate ionization chamber (volume: 0.0053 cm3, type 34013, PTW, Freiburg, Germany) and a dosimeter (UNIDOS E, PTW). Two waterproof measuring chambers in the water tank were designed to measure depth dose rate and isotropy respectively. The ionization chamber(IC) is inserted with the ionization holder into the measuring chamber, which is closed with cover during measurement. “dIC” is designated as “distance between the entrance foil and chamber top” in PTW calibration certificate. The thickness of measuring chamber was printed in the user manual as “dH”. The air gap “dA” between the upper surface of ionization chamber housing and the inside of measuring chamber wall is constantly 0.5 mm. Those distances are considered in measuring the depth dose rate at particular position. The XRS is fixed into the platform which can be adjusted vertically with range of 10 cm and precision of 0.001 mm. The lowest surface of the probe was defined as the isocenter of 0 mm in depth. Due to the certain thickness of the holder wall of the ionization chamber, the measurement in Z direction ranges from 2.0 mm to 44.0 mm with 0.5 mm increment when the distance is less than 40.0 mm, 1.0 mm for 40.0 mm or higher with a period of 1 minute for each point of measurement. The platform of the water tank is also designed with a turntable structure in the X/Y plane for measuring the isotropic dose distribution. The XRS can be rotated with the platform every 45˚ to the total of 8 angles (initial position was defined as 0˚ during measurements). The X, Y directions can be located by fine tune to align the central axis of the bare source probe with the center of the ionization chamber. The position of X, Y directions were kept unchanged for isotropic measurement.
……
3.1. Depth Dose Rate of Bare Probe The depth dose rate for the bare probe was measured three times. Figure 2 depicts the dose rate in Gy/min or error of measurement in percentage as a function of distance between the surface of XRS probe and the ionization chamber monitor-node. The red line represents the average measurement for three times, the data in blue line are obtained from the operation system and the repeatability of the measurements for 3 times are shown in percentage of error in black line. As shown in Figure 2(a) & Figure 2(b), the maximum deviation of three times measurement for bare probe ranges from 0.14% to 1.3%. The value of dose rate measured is higher than that obtained from the operation system when the distance between the surface of XRS probe to the ionization chamber monitor-node is <5.0 mm; the value measured is lower than that from the operation system when the distance is 10.0 mm or more. The average error between the actual measurement and the system is −2.16% ± 1.36% ranging from −3.65% to 2.83%.
3.2. Isotropy of Bare Probe Figure 3(a) depicts isotropic dose in X/Y plane for bare probe, the value of each angle was normalized to the average of 8 measurements at the specific distance. The consistency for the value of each measurement angle is relatively good. As shown in Figure 3(b), the relative deviation of measurements for four times ranged from −1.9% to 2.1%. The deviation of measurement at 90˚ (−1.9%) and 225˚ (2.1%) were relatively larger. The tendency of the measuring error is consistent with good repeatability.
Figure 1. Schematic diagram of commissioning setup for a spherical applicator of 4.5 cm in diameter (water tank).
Figure 2. The depth dose rate (a) and the error in percentage; (b) measured in comparison with data from operation system for bare probe. The red line represents the average of measurement for three times, blue line depicts the data from the operation system, the black line for the deviation of measurement with bare probe in percentage error.
Figure 3. Isotrophy in X/Y plane for bare probe. (a) Normalized value and relative deviation in 8 measurement angles. The hollow circle represents the normalized value, and solid dots the relative percentage deviation of each measurement point magnified 20 times; (b) Relative percentage deviation for 4 times measurement.
3.3. Depth Dose Rate of Spherical Applicators Figure 4(a) depicts the curve of dose rate measured as a function of the distance from the surface of spherical applicators to ionization chamber monitor-node for 8 types of spherical applicators. The dose rate is the highest at the surface of the applicators which falls as the distance between the surfaces of spherical applicators to ionization chamber monitor-node increases. The dose rate varies significantly with the size of the applicator. For the applicator with smaller diameter the dose rate is higher with relative shorter treatment time if the same radiation dose is delivered. As the diameter of the applicator increases, the falling gradient in dose rate slows down. From Figure 4(a), we also observe that the curves of the dose rate overlaps for applicators in diameters 2.5 cm and 3.5 cm, 3.0 cm and 4.0 cm, respectively.
Figure 4(b) shows the curve of depth dose rate as a function of distance between the isocenter of spherical applicator and ionization chamber monitor-node. Obvious separation of dose rate was observed. At the same distance from the isocenter, the dose rate is smaller in the applicators of diameter ≤3.0 cm than those of diameter >3.0 cm. The difference in depth dose rate, transfer coefficient and isotropy between measurements and system data is presented in Table 1. The average deviation in depth dose rate between measurements and system data for the applicators of diameter ≤3.0 cm is in the range from −5.1% to −4.8%; it reduces to −2% – −0.1% with the increase in diameter of the applicators.
Figure 4. The curve of depth dose rate measured for each spherical applicator. (a) The curve of depth dose rate of different distance from the surface of spherical applicators to ionization chamber monitor-node; (b) The curve of depth dose rate of different distance from the isocenter of spherical applicator to ionization chamber monitor-node.
Table 1. The difference in depth dose rate, transfer coefficient and isotropy for spherical applicators.
Diameters (cm)   ……
Deviation of depth dose rate (range) (%) ……
Deviation of transfer coefficiency (range) (%) …….
Isotropy (range) (%) …….
3.4. Transfer Coefficient of Spherical Applicators The transfer coefficient of spherical applicator is defined as ratio between the depth dose rates with or without applicator at the same distance to the source isocenter. The depth dose rate can be obtained by multiplication of the bare source dose rate and the conversion coefficient. Figure 5 is the curve for deviation in transfer coefficient between measured values and the system data for spherical applicators. The average deviation of transfer coefficient at the same distance between the isocenter of spherical applicator and ionization chamber monitor-node for applicators ≤3.0 cm ranges from −2.6% to −2.2%; whereas the average deviation for tors >3.0 cm is between 0.8% and 2.4%. The more detailed data for difference in transfer coefficient between measurements and system data are presented in Table 1 for each individualized applicator.
3.5. Isotropy of Spherical Applicator As shown in Table 1, the average deviation in isotropy between measurements and system data of different spherical applicator’s X/Y plane ranges from −1.4% to 2.6% which shows no obvious change as the diameter of applicator increases. Figure 6(a) & Figure 6(b) depicts the representative isotropy for dose distribution in X/Y plane for a spherical applicator of diameter 4.5 cm which shows the consistency of the measurement values.
Figure 5. The curve for deviation in transfer coefficient between measured values and the system data for spherical applicators.
Figure 6. The representative isotropy in X/Y plane for a ϕ 4.5 cm spherical applicator. (a) Point view of normalized value and the relative deviation; (b) The percentage relative deviation of 3 times measurements.
3.6. Repeatability of Spherical Applicator The Dose rate was measured in 3 times for 4.5 cm spherical applicator, the deviation in repeatability ranges from 0.2% to 0.7%, which shows good repeatability and it gradually gets better as the depth increases. As shown in Figure 7(a) & Figure 7(b), poorer repeatability and increased deviation of error are observed when the distance >15 mm from the surface.
Figure 7. The curve for measurement value of depth dose rate and comparison chart with system data for ϕ 4.5 cm spherical applicator. (a) The depth dose rate for ϕ 4.5 cm spherical applicator. The red line intends for the average 3 measurement values, the blue line for the system value; (b)The error in percentage for the measurements in 3 times.
3.7. Potential Limitation of Clinical Application Figure 8 depicts the graphic of dose distribution for the spherical applicator in diameter 4.5 cm. The doses measured at the distance of 0.2, 0.5, 1.0, and 2.0 cm are 15.3, 10.7, 6.4 and 2.7 Gy, respectively when surface dose of 20 Gy is prescribed. Table 2 lists the depth dose value for spherical applicators when the surface dose of 20 Gy is prescribed, which shows fast dose falling with gradual increasing depth. The doses range from 35.0% to 56.0% at 5 mm from the surface of the spherical applicators and fall to 16.5% – 34.0% at 10 mm and 5.5% – 15.5% at 20 mm.
Figure 8. Dose distribution for a ϕ 4.5 cm spherical applicator when the surface dose of 20 Gy is prescribed.
Table 2. Depth dose rate (Gy) at serial typical depths for spherical applicators when surface dose of 20 Gy is prescribed.
Diameter (cm)
Distance from surface (mm) 0 2 5 10 15 20
Dose measurement tools used for INTRABEAM system include water tank/ionization chamber [10] [11], film/ solid water phantom [11], Thermo-luminescence [12], etc. Water tank/ionization chamber has the highest accuracyamong those instruments [13]. Schneider et al. have compared homogeneity of each dose distribution and depth-dose measurements for flat and surface applicators using film dosimetry in a solid water phantomand a soft X-ray ionization chamber in a water tank [10]. One of the important factors influencing the accuracy of the measurement is the relative position error of X-ray source to ionization chamber due to sharp attenuation of low energy X-ray in the water. The minimal position change will lead to relatively large measurement deviation. As shown in depth dose rate curve fitting function and differential coefficient data, the dose gradient at 3 and 10 mm from the source isocenteris 60%/mm and 24%/mm, respectively. The position error of ±0.1 mm will lead to deviation of dose rate at ±6% and ±2.4%, respectively.
Eaton et al. have reported that the positioning of X-ray source is the most important factor that affects the measurement result and the measurement accuracy may be affected by the ionization chamber’s volume effects in the area of steep dose falling, which may account for the poorer repeatability of measurement near the isocenter [13]. Other factors influence the accurate measurement including the energy response of ionization chamber [13], low voltage of X-ray [14], etc. The spectrum of X-ray approaching to the surface of the probe is complex ranging from 0 to 50 kV with a large number of low energy kV X-ray [14] (<20 kV) which showed rapid attenuation in water [15] resulting in significant difference at different depth and the measurement deviation could reach to ±2.2% due to the energy response [9].
Another important reason for poor repeatability is that the measurement can be affected by noise current. The sensitive volume of ionization chamber is only 0.0053 cm3, and the ionization chamber is supposed to collect only approximately 25 pC charge per minute at the depth of 35 mm during the measurement of X-ray source, which has led to low signal-to-noise ratio influenced by noise current. The uncertainty of measurement for ionization chamber could reach to ±3.4% [9]. The accuracy and repeatability of measurement will gradually decrease with further increase in the distance from the X-ray source since ionization chamber collects less charge per minute (about 16.5 pC).
The type of ionization chamber, design of measuring chamber for water tank, method for calculation and condition of acquisition system for absorbed dose are different between users and company system. The calculation of absorbed doses, transfer coefficient and isotropy in X/Y plane for the measurement data requires the manufacturer to provide the correction function f'(R) for bare probes and spherical applicators of different size. The f'(R) value is different at different depth which may introduce correction error, the error for uncertainty of f'(R) is relatively larger when the depth is shallower. The average of uncertainty for f'(R) is ±7.8%.
Armoogum et al. analyzed the factors that affect the measurement result including temperature and atmospheric pressure correction factor, ionization chamber position deviation, ionization chamber current, the chamber/dosimeter calibration factor, output drift and calibration of absorbed dose [9]. Among these factors positional deviation is the most significant one. The estimation of total uncertainty for all these factors can reach to ±10.8%. Our data have shown that deviation between the measurement data and the system ranges from −10.0% to −5.0% for spherical applicator ≤3 cm in diameter and within ±5.0% for the applicators>3 cm in diameter, which are comparable with the data reported in literatures [9].
Our data have also shown a relative larger deviation of isotropy in X/Y plane measured at 90˚ (1.9%) and 225˚ (2.1%). The possible reasons include: 1) the probe is a hollow needle with 100 mm long and 3.2 mm outside diameter which may be bended during the operation with need to be calibrated before each use since system requirements of probe bending value is less than 0.1 mm; 2) there may exist certain errors during the rotation of the water tank platform, which may lead to different distance from the tip of the probe to the ionization chamber monitor units in different angles; 3) the noise current may influence the correct measurements (the dosimeter charge is about 16.5 pC/min without applicators).
The clinical application of IORT with INTRABEAM system is determined by the size and category of applicators. The biggest diameter for spherical and tablet applicators is 5.0 cm and 6.0 cm, respectively. The surrounding wall is embedded with thin metal sheath interiorly for spherical applicators of diameter ≤3 cm. Low energy X-ray can be attenuated rapidly through the metal sheath. The gradient of dose rate value for spherical applicators in diameter ≤3 cm were larger than that for applicators with diameter >3cm when measured at the same distance from the isocenter.
Our results have shown that the steep dose gradient exists from 0 cm, the surface of the applicator to 1.0 cm; the higher surface dose rate and the greater dose gradient have been observed for the smaller applicators. When the region of treatment is too large, it will be very difficult to calculate accurate dose distribution due to lack of beam bridging technology with over- or under-dose. The limited region and depth of treatment may restrain from its potential in clinical application. Vaidya JS et al. have reported in TARGIT A, a phase III clinical trial that for patients with early breast cancer <2 cm the tumor local control rate of IORT is not inferior to external beam radiotherapy after breast conserving surgery; external beam radiotherapy should be considered for patients with tumor of 2 – 3 cm or with poorer prognostic factors [6]. Sperk E et al. have reported that patient selection for IORT should be restrictive when provided as accelerated partial breast irradiation (APBI) [4].
The system only uses water for dose calculation whereas homogeneity of human structure cannot be revised. The calculated dose value and the deviation will be very big in different tissue due to the spectrum chacteristics of the low energy photon (≤50 kV); the dose distribution will be influenced by the air gap between tissues and the applicator [14]. Monte Carlo modelling could be used in comparison with result from ionization chambers, radiochromic film and other dosimeters such as TLDs on the subsequent periodic QA tests for the INTRABEAM system [8]. There exist great difficulties for external beam radiotherapy when needed to supplement the postoperative radiotherapy especially for the accurate dose calculation and precise delineation of target and surrounding normal tissue irradiated especially for the nerves and blood vessels. All these factors mentioned above may seriously restrict INTRABEAM system to be widely used in clinical practice.
5. Conclusion In summary, thorough commissioning of INTRABEAM system helps us better understand the dosimetry characteristics, verify the system data and a quire adequate data for clinical application and routine quality assurance. It is necessary to establish the benchmark for long term quality assurance based on the measurement data. The characteristics of high dose at the surface of applicator, great dose gradient and limited treatment range may restrain from its potential in wide clinical application

References

[1] Beddar, A.S. (2005) Stability of a Mobile Electron Linear Accelerator System for Intraoperative Radiation Therapy. Medical Physics, 32, 3128-3131.   http://dx.doi.org/10.1118/1.2044432
[2] Klink, C.D., Binnebosel, M., Holy, R., Neumann, U.P. and Junge, K. (2014) Influence of Intraoperative Radiotherapy (IORT) on Perioperative Outcome after Surgical Resection of Rectal Cancer. World Journal of Surgery, 38, 992-996.    http://dx.doi.org/10.1007/s00268-013-2313-1
[3] Goubert, M. and Parent, L. (2015) Dosimetric Characterization of INTRABEAM((R)) Miniature Accelerator Flat and Surface Applicators for Dermatologic Applications. Physica Medica, 31, 224-232. http://dx.doi.org/10.1016/j.ejmp.2015.01.009
[4] Sperk, E., Astor, D., Keller, A., Welzel, G., Gerhardt, A., Tuschy, B., et al. (2014) A Cohort Analysis to Identify Eligible Patients for Intraoperative Radiotherapy (IORT) of Early Breast Cancer. Radiation Oncology, 9, 154.
http://dx.doi.org/10.1186/1748-717X-9-154
[5] Sedlmayer, F., Reitsamer, R., Fussl, C., Ziegler, I., Zehentmayr, F., Deutschmann, H., et al. (2014) Boost IORT in Breast Cancer: Body of Evidence. International Journal of Breast Cancer, 2014, Article ID: 472516.
http://dx.doi.org/10.1155/2014/472516
[6] Vaidya, J.S., Wenz, F., Bulsara, M., Tobias, J.S., Joseph, D.J., Keshtgar, M., et al. (2014) Risk-Adapted Targeted Intraoperative Radiotherapy versus Whole-Breast Radiotherapy for Breast Cancer: 5-Year Results for Local Control and Overall Survival from the TARGIT—A Randomised Trial. Lancet, 383, 603-613.    http://dx.doi.org/10.1016/S0140-6736(13)61950-9
[7] Williams, N.R., Pigott, K.H., Brew-Graves, C. and Keshtgar, M.R. (2014) Intraoperative Radiotherapy for Breast Cancer. Gland Surgery, 3, 109-119.
[8] Eaton, D.J. and Duck, S. (2010) Dosimetry Measurements with an Intra-Operative X-Ray Device. Physics in Medicine and Biology, 55, N359-N369.    http://dx.doi.org/10.1088/0031-9155/55/12/N02
[9] Armoogum, K.S., Parry, J.M., Souliman, S.K., Sutton, D.G. and Mackay, C.D. (2007) Functional Intercomparison of Intraoperative Radiotherapy Equipment—Photon Radiosurgery System. Radiation Oncology, 2, 11.
http://dx.doi.org/10.1186/1748-717X-2-11
[10] Schneider, F., Clausen, S., Tholking, J., Wenz, F. and Abo-Madyan, Y. (2014) A Novel Approach for Superficial Intraoperative Radiotherapy (IORT) Using a 50 kV X-Ray Source: A Technical and Case Report. Journal of Applied Clinical Medical Physics, 15, 4502.
[11] Ebert, M.A., Asad, A.H. and Siddiqui, S.A. (2009) Suitability of Radiochromic Films for Dosimetry of Very-Low Energy X-Rays. Journal of Applied Clinical Medical Physics, 10, 2957.    http://dx.doi.org/10.1120/jacmp.v10i4.2957
[12] Soares, C., Drupieski, C., Wingert, B., Pritchett, G., Pagonis, V., O’brien, M., et al. (2006) Absorbed Dose Measurements of a Handheld 50 kVP X-Ray Source in Water with Thermoluminescence Dosemeters. Radiation Protection Dosimetry, 120, 78-82.    http://dx.doi.org/10.1093/rpd/nci622